Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–263
The power-on default for SRER is 0 if *PSC is 1. If *PSC is 0, the SRER
maintains its value through a power cycle.
*SRE 48
sets the bits in the SRER to 00110000 binary.
*SRE?
might return a value of 32, showing that the bits in the SRER have the binary
value 00100000.
*STB? (Query Only)
(Read Status Byte) query returns the contents of the Status Byte Register (SBR)
using the Master Summary Status (MSS) bit. For a complete discussion of the
use of these registers, see page 3–1.
Status and Error
*CLS, DESE, *ESE, *ESR?, EVENT?, EVMSg?, FACtory, *SRE
*STB?
*STB
?
<NR1>
*STB?
might return the value 96, showing that the SBR contains the binary value
01100000.
TEKSecure
Initializes both waveform and setup memories. This overwrites any previously
stored data.
TEKSecure writes zeros in all waveform reference memory, regardless of
selected record length, and puts all setups in the factory init state.
TEKSecure then verifies that the waveform and setup memory are in the desired
state. It asserts a pass or a fail event on completion.
Examples
Group
Related Commands
Syntax
Returns
Examples