Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–169
HORizontal:FASTframe:COUNt?
::
FASTframe
HORizontal
COUNt
?
<Space> <NR1>
<NR1> indicates the number of frames to acquire.
HORIZONTAL:FASTFRAME:COUNT 2
Sets up FastFrame mode to acquire two frames (segments) of data.
HORizontal:FASTframe:LENgth
TDS 500C & 700A Only
Setup length of each FastFrame frame. This is equivalent to setting FastFrame
Setup in the Horizontal menu and the Frame Length menu item in the side
menu. FastFrame, also known as memory segmentation, lets users capture a
series of triggered acquisitions with minimal, intervening, time between them.
Horizontal
HORizontal:FASTframe:LENgth <NR1>
HORizontal:FASTframe:LENgth?
::
FASTframe
HORizontal
LENgth
?
<Space> <NR1>
<NR1> indicates the frame (segment) length.
HORIZONTAL:FASTFRAME:LENGTH 250
Sets up the FastFrame mode to acquire frames of 250 samples each.
HORizontal:FASTframe:POSition
TDS 500C & 700A Only
Display the selected FastFrame frame. This is equivalent to selecting Horiz Pos
in the Horizontal menu, selecting the Frame menu item in the side menu, and
entering a value with the keypad or the general purpose knob. FastFrame, also
Arguments
Examples
Group
Syntax
Arguments
Examples