Agilent Technologies E4350B Landscape Lighting User Manual


 
Service Addendum 7
Table 2-15B. Performance Test Record for Agilent Model E4351B
Agilent E4351B Report No.______________ Date _________________
Test Description Minimum Spec. Result * Maximum Spec. Measurement
Uncertainty
Constant Voltage Tests
Voltage Programming and
Readback Accuracy
Low Voltage (set to 0V) V
out1
Readback Accuracy
High Voltage (set to 120V) V
out2
Readback Accuracy
Load Effect
Source Effect
PARD (Ripple & Noise)
Peak to Peak
RMS
-20mV
V
out1
– 84mV
119.89V
V
out2
– 180mV
– 2mV
– 2mV
0
0
_________ mV
_________ mV
__________ V
_________ mV
_________ mV
_________ mV
_________ mV
_________ mV
+20mV
V
out1
+ 84mV
120.11V
V
out2
+ 180mV
+ 2mV
+ 2mV
95mV
12mV
1.7µV
1.7µV
1.7mV
1.7mV
230µV
230µV
3mV
1.5mV
Constant Current Tests
Current Programming and
Readback Accuracy
Low Current (set to 0A) Iout1
Readback Accuracy
High Current (set to 4A) Iout2
Readback Accuracy
Current Sink (0.9A) Readback
Accuracy
PARD (Ripple & Noise)
RMS
SAS Mode **
RMS
Peak to Peak
Load Effect
Source Effect
– 10mA
I
out1
– 14mA
3.982A
I
out2
– 22mA
Isink – 27mA
0
0
0
– 1mA
– 1mA
_________ mA
_________ mA
__________ A
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
+ 10mA
V
out1
+ 14mA
4.018A
V
out2
+ 22mA
Isink + 27mA
2.4mA
4mA
32mA
+ 1mA
+ 1mA
16µA
16µA
0.6mA
0.6mA
120µA
270µA
270µA
2.6mA
2.2µA
2.2µA
* Enter your results in this column
** PARD (Optional Test in SAS mode)
Program I – V Curve: I
SC
= 4A I
MP
= 3.75A V
MP
= 120V
V
OC
= 130V
R
L
= 32 Ohms.
Use Fig 2-3 with scope across monitoring resistor for P-P ripple or RMS voltmeter for RMS.