Agilent Technologies E4350B Landscape Lighting User Manual


 
Service Addendum6
Performance Test Record
Table 2-15A. Performance Test Record for Agilent Model E4350B
Agilent E4350B Report No.______________ Date _________________
Test Description Minimum Spec. Result * Maximum Spec. Measurement
Uncertainty
Constant Voltage Tests
Voltage Programming and
Readback Accuracy
Low Voltage (set to 0V) V
out1
Readback Accuracy
High Voltage (set to 60V) V
out2
Readback Accuracy
Load Effect
Source Effect
PARD (Ripple & Noise)
Peak to Peak
RMS
-10mV
V
out1
– 42mV
59.945V
V
out2
– 90mV
– 2mV
– 2mV
0
0
_________ mV
_________ mV
__________ V
_________ mV
_________ mV
_________ mV
_________ mV
_________ mV
+10mV
V
out1
+ 42mV
60.055V
V
out2
+ 90mV
+ 2mV
+ 2mV
95mV
12mV
1.5µV
2µV
845µV
845µV
40µV
40µV
840µV
2.1mV
Constant Current Tests
Current Programming and
Readback Accuracy
Low Current (set to 0A) Iout1
Readback Accuracy
High Current (set to 8A) Iout2
Readback Accuracy
Current Sink (Isink = 1.8A)
Readback Accuracy
PARD (Ripple & Noise)
RMS
SAS Mode **
RMS
Peak to Peak
Load Effect
Source Effect
– 20mA
I
out1
– 28mA
7.956A
I
out2
– 30mA
Isink – 48mA
0
0
0
– 1mA
– 1mA
_________ mA
_________ mA
__________ A
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
_________ mA
+20mA
V
out1
+ 28mA
8.044A
V
out2
+ 30mA
Isink + 48mA
2.4mA
4mA
32mA
+ 1mA
+ 1mA
0.5µA
20µA
3.3mA
3.3mA
1mA
1.8mA
270µA
2.6mA
3.4µA
3.4µA
* Enter your results in this column
** PARD (Optional Test in SAS mode)
Program I – V Curve: I
SC
= 8A I
MP
= 7.5A V
MP
= 60V
V
OC
= 65V
R
L
= 8 Ohms.
Use Fig 2-3 with scope across monitoring resistor for P-P ripple or RMS voltmeter for RMS.