Agilent Technologies E4350B Landscape Lighting User Manual


 
Service Addendum 5
Voltage and Current Values
Table 2-3. Voltage and Current Values
Agilent Model Full-scale
Voltage
Max. Prog.
Voltage
Full-scale
Current
Max. Prog.
Current
Max. Prog.
Overvoltage
E4350B
60V 61.5V 8A 8.16A 73.0V
E4351B
120V 123V 4A 4.08A 140V
Transient Recovery Time
This test does not need to be performed on the Agilent E4350B.
CC RMS Noise Measurement Test Setup
Figure 2-3. CC RMS Noise Measurement Test Setup
+
-
+
IM
-
IM
IP
+
S
-
S
+240 VDC MAX
-
-
I
VP I P
SET TO
REMOTE
REMOTE
LOCAL
VOLTMETER
RMS
MONITOR
RESISTOR 1.6 OHM
(Adjusted to 1 Ohm +/- 1%)
LOAD RESISTOR
E4350B 2 16 Ohm resistors in parallel
(Adjusted to 7 Ohms +/- 1%)
E4351B 2 16 Ohm resistors in series
(Adjusted to 25 Ohms +/- 1%)