Communication format Format: OFF, MODE 1, MODE 2, MODE 3, MODE 4,
(FORMAT) STD 1 (For shipment, “OFF” is set.)
MODE 1:When
is pressed, communication is done
after printing.
MODE 2:Data communication is done every measure-
ment.
MODE 3:When
is pressed, communication is done
without printing.
MODE 4:When
is pressed, communication is done
without printing.
STD 1: When
is pressed, communication is done
after printing.
Communication speed (SPEED) Baud rate: 2400, 9600 (bps) (For shipment,
2400bps is set.)
51
REFERENCE
The Topcon CT-80 complies with the CE marking.
Before connecting a personal computer to the TOPCON
product, make sure that such external equipment is in
compliance with the CE marking.
Under the RTS-CTS control, if no CS (CTS) signal is returned from the receiver side,
transmission can be canceled by pressing
. Also, if no DR (DSR) signal is re-
turned, it is recognized as a communication failure and FAIL is displayed on the
monitor screen.
When Mode 1, 2 or 3 is set, data is transmitted automatically without confirmation
from the receiver side. When STD 1 is set, communication is controlled by RTS-CTS.