Topcon COMPUTERIZED TONOMETER Multi-tool User Manual


 
3
Hold the paper and pull it diagonally to cut.
HOW TO CORRECT MEASUREMENT VALUES
Though the machine is adjusted for displaying optimal measurement values, the values
can be corrected within a -4 +3mmHg range.
1
Make sure the power is OFF.
2
Open the control panel lid.
3
Using a screwdriver, turn the “ ” of the select switch.
Settings: 0: Base setting For shipment, the switch is set here.
1: +1mmHg Base setting + 1mmHg
2: +2mmHg Base setting + 2mmHg
3: +3mmHg Base setting + 3mmHg
F: -1mmHg Base setting - 1mmHg
E: -2mmHg Base setting - 2mmHg
D: -3mmHg Base setting - 3mmHg
C: -4mmHg Base setting - 4mmHg
4
Turn the
POWER SWITCH
ON.
The measurement procedure is exactly the same as with the correction.
43
INDIVIDUAL OPERATIONS
To avoid paper jams, cut the paper carefully and evenly.
MEMO
Never set the select switch for more than eight points to avoid
malfunctions.
MEMO
Select switch