Topcon COMPUTERIZED TONOMETER Multi-tool User Manual


 
SAFETY CAUTIONS
Icon Prevention item Page
To avoid potential injury, hold the instrument in the proper position.
13
To avoid electrical shock, do not handle the power plug with wet
fingers.
14
Never insert your fingers under the measuring head.
Inform the patient of this, too.
Careless insertion of fingers may cause injury by pinching.
30
Never insert your fingers under the measuring head.
Inform the patient of this, too.
Careless insertion of fingers may cause injury by pinching.
34
Do not use or apply any spray-typed cleaner near the instrument.
If a drop of cleaner remains inside the measuring nozzle, the pa-
tient’s eye may be injured during measurement.
56
Before carrying the instrument, be sure to affix it firmly by turning
the fixing screw at the base.
If the instrument is moved with the screw loosened, it may result in
damage to the instrument.
13
When moving the instrument, be sure to hold it at the bottom sur-
face with two people. Carrying by one person may cause back inju-
ry or injury by falling parts. Also, holding areas other than the bot-
tom surface may cause pinching fingers between parts and injury
by falling parts as well as damage to the instrument.
13
Before measuring, set the safety stopper. If the safety stopper is
not set, it may cause injury to the eye that comes in contact with the
measuring window glass. Set the safety stopper separately for the
right and left eyes.
31
When setting the safety stopper, do it from the instrument side
(safety stopper knob side).
Setting from another position does not easily allow you to check the
positions of the measuring window glass and the patient’s eye and
may cause injury to the eye that comes in contact with the measur-
ing window glass.
31
To clean the measuring window glass and the window glass inside
the measuring nozzle, use ethanol. Using other chemicals may
cause damage to the patient's eye during measurement.
52
53
CAUTION
4