Fluke PM6685R Outdoor Timer User Manual


 
Command Reference 9-123
*IDN?PM6680B/81/85
Identification query
Reads out the manufacturer, model, serial number, Firmware level for main and
GPIB program in an ASCii response data element. The query must be the last
query in a program message.
Response is <Manufacturer> , <Model> , <Serial Number>, <Firmware Level>.
<Serial number> is not implemented in PM6680B and PM6685 and will always re
-
turn a zero. Please look at the type plate at the rear panel of the counter if you are
interested in the serial number. PM6681 returns the correct serial number.
Example
SEND ®*IDN?
READ¬ Fluke, 8888PM6685, 8 0, 8 MAIN 8 V1.01 8 19 Nov 8
1992 8 / 8 GPIB 8 V1.12 8828 8 Oct 8 1992
Complies to standards: IEEE 488.2 1987.
*LMC?PM6680B/81/85
Learn Macro
Makes the instrument send a list of string data elements, containing all macro la-
bels defined in the instrument.
Returned Format:
<String> { ,<String> }¿
<String> = a Macro label. (String data will be surrounded by“”asintheexample below.)
Example:
SEND® *LMC?
May give the following response:
READ
¬
“AUTOFILT”,"AMPLITUDE?"
Complies to standards: IEEE 488.2 1987.