9-92 Command Reference
:AVERage :STATe PM6680B/81/85
8 <Boolean>
Average or Single?
Switch on/off the average function.
Parameters: <Boolean>=1|ON/0|OFF
ON means multiple period measurements for period related measurements and time-interval av
-
erage for Time-Interval measurements.
OFF means that the counter measures on a single cycle. This is the same as when pressing the
SINGLE key on the front panel.
When Single is selected and an array measurement is done, the Measurement
Time, set by :Acquisition:APERture, sets the time between the measure
-
ments in the array. This means that if you want a very high speed yo must set
:AVER:STATE 8 OFF and :ACQ:APER 8 MIN.
Returned format: <Boolean>¿
*RST condition: ON
:FREQuency :RANGe :LOWer PM6680B/PM6681
8 «<Numeric value>|MIN|MAX»
High Speed Voltage Measurements
Use this command to speed up voltage measurements and Autotrigger functions
when you don’t need to measure on low frequencies.
Time to determine trigger levels
Measuring
function
Min frequency limit (Default) Max frequency limit
PM6680B PM6681 PM6680B PM6681
Freq A
48 ms 85 ms 26 ms 30 ms
Time A-B 82 ms 38 ms
Parameters:
<Numeric value> for PM6680B, 100 gives the lower frequency limit of 100 Hz
and 10000 for a lower frequency limit of 10 kHz.
MIN gives 100 Hz frequency limit for PM6680B and 1Hz for PM6681.
MAX gives 10 kHz frequency limit forPM6680B and 50 kHz for PM6681.
Returned format:
<Numeric value>¿
*RST condition: 100
Complies to standards: SCPI 1991.0, confirmed.