Fluke PM6685R Outdoor Timer User Manual


 
Command Reference 9-59
:MEASure_«:DCYCle/:PDUTycycle»PM6680B/81/85
[8 [<threshold>] [,(@«1|2|4|6»)]]
Positive Duty Cycle
Traditional duty cycle measurement is performed. That is, the ratio between the
on time and the off time of the input pulse is measured.
Parameters
<threshold> parameter sets the trigger levels in volts. If omitted, the auto trigger level is set to
50 percent of the signal.
(@«1|2|4|6») is the channel to measure on:
(@1) means input A
(@2) means input B (Only PM6680B and PM6681)
(@4) means input E (Rear panel arming input)
(@6) means the internal reference
If you omit the channel, the instrument measures on input A (@1).
Example:
SEND® :MEAS:PDUT?
READ¬ +5.097555E-001
In this example, the duty cycle is 50.97%
Complies to standards: SCPI 1991.0, confirmed.
EXPLANATIONS OF THE MEASURING
FUNCTIONS
This sub-chapter explains the various measurements that can be
done with :MEASure and :CONFigure;:READ. Only the queries
for single measurements using the measure command are given
here, but all of the information is also valid for the :CONFigure
command and for both scalar (single) and array measurements.