Fluke PM6685R Outdoor Timer User Manual


 
Command Reference 9-67
:MEASure :PERiod?PM6680B/81/85
[8 [<expected value> [,<resolution>]][,<(@«1|2|3|4|5|6|7»)>]]
Period
A traditional period measurement is performed.
The <expected value> and <resolution> are used to calculate the Measurement
Time ([:SENSe]:ACQuisition:APERture).
Parameters:
<expected value> is the expected Period,
<resolution> is the required resolution,
<(@«1|2|3|4|5|6»)> is the channel to measure on:
(@1) means input A
(@2) means input B (Only PM6680B and PM6681)
(@3) means input C (HF-input option)
(@4) means input E (Rear panel arming input)
(@5) means input A prescaled by 2
(@6) means the internal reference
(@7) means input A with the variable hysteresis mode (Only PM6680B)PM6681)
If you omit the channel, the instrument measures on input A (@1).
Complies to standards: SCPI 1991.0, confirmed.
:MEASure :PHASe?PM6680B PM6681
[8 [<expected value>[,<resolution>]] [,(@«1|2»),(@«1|2»)]]
Phase
A traditional PHASe measurement is performed.
Parameters:
<expected value> and <resolution> are ignored by the counter
The first (@«1|2») is the start channel and the second (@«1|2») is the stop channel
(@1) means input A
(@2) means input B
If you omit the channel, the instrument measures between input A and input B.
Complies to standards: SCPI 1991-0, approved.