Fluke 196C Telescope User Manual


 
Using The Recorder Functions
Recording Scope Waveforms In Deep Memory (Scope Record)
3
35
Using External Triggering to Start or Stop Scope
Record
To record an electrical event that causes a fault, it might
be useful to start or stop recording on an external trigger
signal:
Start on Trigger to start recording; recording stops when
the deep memory is full
Stop on Trigger to stop recording.
Run When Triggered to continue recording as long as a
next trigger comes within 1 division in view all mode.
To set up the test tool, continue from step 3 of the
previous section:
4 Apply the signal to be recorded to the red BNC
input A. Apply a trigger signal to the red and
black external trigger banana inputs. (See
Figure 19.)
5 Open the Recorder Options
menu.
Figure 19. Scope Record Using External Triggering
6 Jump to Display Glitches:.
7 Jump to Mode:.
8 Select on EXT. ... to open the
Single Sweep on Ext. menu.
9 Select one of the Conditions:
and jump to Slope:.