Fluke 196C/199C
Users Manual
94
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges.............................same as VDC, VAC, VAC+DC
Probe Sensitivity.................100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy..........................same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)
Peak
Modes...........................Max peak, Min peak, or pk-to-pk
Maximum Voltage
with 10:1 probe...................................................1000 V
direct (1:1)............................................................300 V
Maximum Resolution
with 10:1 probe....................................................10 mV
direct (1:1).............................................................1 mV
Full Scale Reading..........................................800 counts
Accuracy
Max peak or Min peak...............................±0.2 division
Peak-to-peak............................................. ±0.4 division
Frequency (Hz)
Range......................................1.000 Hz to full bandwidth
Full Scale Reading...................................... 9 999 counts
with at least 10 waveform periods on screen.
Accuracy
1 Hz to full bandwidth......................±(0.5 % +2 counts)
Duty Cycle (DUTY)
Range..................................................... 4.0 % to 98.0 %
Pulse Width (PULSE)
Resolution (with GLITCH off).........................1/100 division
Full Scale Reading......................................... 999 counts
Accuracy
1 Hz to full bandwidth......................±(0.5 % +2 counts)