Specifications
10:1 Probe
9
103
Figure 52. Safe Handling: Max. Input Voltage Between
Scope References, and Between
Scope References and Meter Reference
10:1 Probe
Safety
Max. Input Voltage............................1000 V CAT II
600 V CAT III
Max. Floating Voltage
from any terminal to ground.....................1000 V CAT II
600 V CAT III
up to 400 Hz
Electrical specifications
Input Impedance at probe tip 10 MΩ (±2 %)//14 pF (±2 pF)
Capacity Adjustment Range...........................10 to 22 pF
Attenuation at DC (1 MΩ input).......................10 x (±2 %)
Bandwidth (with FLUKE 199C).....DC to 200 MHz (-3 dB)
Environmental
Temperature
Operating.............................0 to 50 °C (32 to 122 °F)
Storage ..........................-20 to +60 °C (-4 to +140 °F)
Altitude
Operating.......................................3 km (10 000 feet)
Storage ........................................12 km (40 000 feet)
Humidity
Operating at 10 to 30 °C (50 to 86 °F)............... 95 %