![](http://pdfasset.owneriq.net/d/ca/dca83be6-7707-4dfd-b92b-35e03570bdc8/dca83be6-7707-4dfd-b92b-35e03570bdc8-bg59.png)
FM Receivers
88
Measurement Procedure:
1 Connect the Receiver as shown.
On the Test Set:
2 Press the PRESET key.
3 Press the RX key.
Using the knob and data entry keys:
4 Set RF Gen Freq to the receiver operating frequency.
5 Set Amplitude to −47 dBm (1 mV).
6 Set AFGen1 To 60% of the Receiver’s specified maximum frequency de-
viation (typically 3 kHz).
7 Set Ext Load R to the test load resistance.
8 Set AC Level meter to measure Watts.
Additional Equipment
Required
Test Load
Special Test Considerations
See "Receiver Test Loads" on page 46.