9 Record the displayed FM Deviation as Pk+.
Calculate the Measurement:
10 Calculate the Deviation Symmetry as follows:
Deviation Symmetry (in percent) =
For example, =
(Pk +) - (Pk -)
(Pk +)
X 100
(3.010) - (2.971)
(3.010
X 100 = 1.29