Measurement Procedure:
1 Connect the Receiver as shown.
On the Test Set:
2 Press the PRESET key.
3 Press the RX key.
Using the knob and data keys:
4 Select the ENCODER screen.
5 Set Mode to CDCSS.
6 Set Standard to CDCSS.
7 Set Code to the Receiver’s CDCSS 3-digit octal code.
8 Set Send Mode to Cont.
9 Set FM Coupling to DC.
10 Set AFGen2 To to 500 Hz.
11 Select Send.
12 verify the Status Flag indicates Sending.
Additional Equipment
Required
Test Load
Special Test Considerations
See "Receiver Test Loads" on page 46.
To Screen
SIGNALING ENCODER (AF GENERATOR 2)
RF GEN
RF ANL
AF ANL
Audio Out
Status:
Sending
Mode
CDCSS
Standard
CDCSS
Send
Data Rate
134.4
bps
Code
143
TOC Time
200.0
ms
Stop
AFGen2 To
5.00
Hz
FM
Bursts
2
Send Mode
Cont
Sending