Agilent 6800 series setup
Set the initial (immediate) voltage to 0 Vrms.
Set the triggered voltage level to 120 Vrms.
Set the frequency to 60 Hz.
Set the peak current limit to maximum.
Set the transient trigger source to BUS.
Set the trigger synchronization source to Phase.
Set the initial synchronization Phase to 40°
(re-program to 50°, 60°, 70°, 80°, and 90°).
Set the acquisition trigger source to TTLTrg.
Set the data acquisition offset to -409 points.
Enable the output of the ac power source/
analyzer.
Initiate the transient trigger.
Initiate the acquisition trigger.
Send a Bus trigger.
Fetch the peak current measurement.
Upon receipt of the trigger, the ac power source/
analyzer output goes to 120 Vrms when the phase
voltage waveform reaches 40°. The data buffer
acquires 409 instantaneous current data points
before the turn-on event and 3687 instantaneous
current data points beginning at the onset of the
turn-on event (for a total buffer of 4096 current
data points) to generate the peak current measure-
ment. To continue the peak inrush current charac-
terization, change the synchronization phase (from
40° to 50°, 60°, 70°, 80°, and 90°) and repeat the
same procedure.