19
Agilent 6800 series setup
• The ac source output is connected to the ac
input of the equipment under test.
• Set the output waveform to sine.
• Set the rms voltage to Fixed mode.
• Set the frequency to Fixed mode.
• Set the voltage to 230 Vrms.
• Set the frequency to 50 Hz.
• Set the measurement window to Rectangular.
• Enable the output of the ac power source/
analyzer.
• Wait until the ac input to the equipment is in
a steadystate condition.
• Send the harmonic current array measurement
command to return 50 harmonic current
amplitudes.
Using the computer:
• Read the harmonic current array into the
computer.
• Compare the measured harmonic current ampli-
tudes of harmonic numbers 2 through 40 to the
IEC 555-2 quasi-stationary standard limits.