Agilent Technologies 41A Sprinkler User Manual


 
3-12
Triggering the Oscilloscope
To use Edge triggering
To use Edge triggering
The Edge trigger type identifies a trigger by looking for a specified slope and
voltage level on a waveform. You can define the trigger source and rising or
falling edge in this menu. The trigger type, source, and level is displayed in the
upper-right corner of the display.
1Press the Edge key in the Trigger section of the front panel to display
the Edge trigger menu.
2 Press the slope softkey ( ) and select either rising edge or falling
edge. This determines whether the trigger will occur on the rising edge
or the falling edge of the input signal.
Slope Selects rising edge ( ) or falling edge ( ) for the selected trigger
source. The selected slope is displayed in the upper-right corner of the display.
3 Select 1, 2, Ext, or Line as the trigger source.
The trigger source can also be set to channel 3 and 4 on the 4-channel
oscilloscope, or can be set to digital channels D15 through D0 on the
mixed-signal oscilloscope. You can choose a channel that is turned off as the
source for the edge trigger. The selected trigger source is displayed in the
upper-right corner of the display next to the slope symbol:
1 through 4 = analog channels
D15 through D0 = digital channels
E = External trigger
L = Line trigger
Slope
Analog channel
source
Digital channel
source
External or Line
source
Trigger level
or threshold
Trigger
source
Slope