Invacare IH6300 Hot Tub User Manual


 
LABEL LOCATION
Supine Bath Tub 6 Part No 1096926
LABEL LOCATION
!
WARNING
ELECTRICAL SHOCK HAZARD. All
disassembly and maintenance of the Supine
Tub MUST be done by a qualified technician,
certified electrician or plumber. Disconnect
electrical supply or turn OFF circuit breaker
before performing any maintenance to the
Supine Tub.
1097132 Rev.A(1) - 02/00
!
DANGER
ELECTRICAL SHOCK HAZARD. For use with hospital
grade GFCI receptacle ONLY. Failure to do so may
result in an electrical shock and/or serious or fatal
injury. DO NOT remove this label.
1097131 Rev.A(1) - 02/00
! DANGER
The National Electrical Code requires
the use of a ground fault circuit
interrupter on all therapeutic pools
and tubs (hydro-therapeutic tanks).
Failure to follow this requirement
could result in serious or fatal injury.
1095610 Rev.A(1) - 02/00
CAUTION
Some state and local plumbing codes require the
installation of a “reduced pressure zone” (RPZ) assembly
on both the incoming hot and cold supply lines to prevent
back flow contamination into the potable water system.
Invacare recommends you check local plumbing code
requirements to determine if RPZs are required on your
tub installation. RPZs are NOT supplied by Invacare and
are to be provided by the customer if required.
P/N 1150759 Rev A
! WARNING
BEFORE using the
Supine Tub READ and
UNDERSTAND the
Owner's Manual for
proper operation and
safety procedures
.
Read and understand ALL
information on Invacare
disinfectant BEFORE use.
Refer toOwner's Manual
for proper disinfecting
procedures.
1095523 Rev.A(1) - 02/00
! WARNING
For disinfecting use ONLY. DO NOT use
for showering.Refer to Owner's Manual
for proper disinfecting procedures.
1095526 Rev.A(1) - 02/00
! WARNING
DO NOT stand in tub. Tub surfaces are
slippery when wet. Serious injury may occur.
1097288 Rev.A(1) - 02/00
! WARNING
DO NOT STAND ON TUB FRAME.
Frame is slippery when wet. Foot injury possible
when adjusting tub height.
1095525 Rev.A(1) - 02/00
! WARNING
BEFORE immersing patient or operating the
shower/massage wand, check that the
water temperature is between 95˚ and 105˚
fahrenheit (35˚ and 41˚ C) to avoid scalding
the patient.
1095524 Rev.A(1) - 02/00
! WARNING
This dispensing system has been
factory tested and calibrated for use
with INVACARE Disinfectant. The use of
any disinfectant with a dilution ratio or
viscosity level different than that of
INVACARE Disinfectant will result in an
improper mixture of disinfectant and
water.This may cause your system to be
improperly disinfected - or in some
instances, not disinfected at all. If you
choose to use a product other than
INVACARE Disinfectant, Invacare will not
be responsible for the proper calibration
or function of your disinfectant system.
1095611 Rev.A(1) - 02/00
P/N 1150694 Rev A
DO NOT install a heater in the tub. Installing a heater
will violate safety codes and may result in scalding
injuries to the user.
! WARNING
P/N 1150694 Rev A
1095523
1095526
1097288
1095611
1095524
1095525
1097132
1095610
NOTE:ModelIH6300ADHONLY
1097131
1150759
1150694
Disinfectant
Cabinet
PAR9992
0)*'$7(
,19$&$5(&217,18,1*&$5(*5283
$03$&,7<$1'5$7('0,1,080&
FLUFXLWLQWHUUXSWHU*)&,
:,5,1*',$*5$0
'$1*(55,6.2)(/(&75,&6+2&.&RQQHFWRQO\
86(:,5(6,=('217+(%$6,62)&
)256833/<&211(&7,21
WRDFLUFXLWWKDWLVSURWHFWHGE\DJURXQGIDXOW
(OHFWULTXH3ORPEHULH
&<&/(
*UHHQ*URXQG
(OHFWULFDO3OXPELQJ
157/&
6$
5
8
+\GURPDVVDJH%DWKWXE-
/,67('
5
32:(5
6833/<
*)&,
/
92/769$&
$036
02'(/180%(5
&21'8&725621/<
VXSSO\
FLUFXLW
7,0(5
02725
ZLUHV
LQSRZHU
UHTXLUHG
:KLWH1HXWUDO
%ODFN+RW
86(&233(5
3DUW1
X
PEHU3$5