SECTION 6—TROUBLESHOOTING
Part No 1096926 31 Supine Bath Tub
SECTION 6—TROUBLESHOOTING
NOTE:Ifproblemsarenotremediedbythesuggestedmeansabove,contactanInvacareDealer.
SYSTEM/OPERATION SYMPTOMS FAULTS SOLUTION
Adjustable Height
Model IH6300ADH
Tub will not raise
or lower
Glide pads not greased Grease glide pads.
Refer to Maintenance for
IH6300ADH Adjustable Height
ONLY Glide Pads on page 26.
Bad electrical connections Check electrical connections at
the control box.
Actuator overloaded Wait 5 minutes and try adjusting
height again.
Actuator burned out Contact an Invacare Dealer.
Obstacle in tub path Inspect and remove obstacle.
No electrical power to tub Check and reset breaker or
GFCI.
Hydromassage System
System will not start No electrical power to tub Check and reset breaker or
GFCI.
System not prepared Be sure water level is above jets.
Check that inlet plug on tub is
depressed.
Switch hose disconnected Reconnect hose to switch and
motor.
System is ineffective Water deposits in
whirlpool jets
Disassemble and clean whirlpool
jets.
Whirlpool jets closed Open whirlpool jets.
Aerator closed Open aerator.
Whirlpool intake clogged Inspect intake and remove
obstruction.
Disinfecting System
No discharge from
disinfectant wand
Whirlpool jets open Close whirlpool jets.
Whirlpool intake open Close whirlpool intake.
Hose kinked Inspect hose and unkink.
Disinfectant siphon tube
kinked
Inspect siphon tube and unkink.
Flow meter adjustment
knob turned off
Turn flow meter adjustment knob
to proper setting.
Tub Operation
No water flow Water supply turned off Turn on water supply.
Restricted water flow RADA valve. Refer to
RADA owner’s manual.
Inspect and clean RADA valve.