Invacare IH6300 Hot Tub User Manual


 
SECTION 2—DISINFECTING THE TUB
Supine Bath Tub 12 Part No 1096926
SECTION 2—DISINFECTING THE TUB
ƽ WARNING
Disinfect the tub after EACH use to avoid resident infection and contamination of the tub.
DO NOT disinfect the tub when it is occupied otherwise injury may occur.
Read and understand all information on Invacare brand disinfectant before use.
ALWAYS wear rubber gloves, an apron and a face shield when using disinfectant. In case of eye or skin
contact, follow procedures on disinfectant container.
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare disinfectant.
The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level different than that of
Invacare disinfectant will result in an improper mixture of disinfectant and water. This may cause your system
to be improperly disinfected - or in some instances, not disinfected at all. If using any disinfectant with a dilution
ratio or viscosity level different than that of Invacare disinfectant or if switching from one type of disinfectant to
another type, the disinfectant flowmeter MUST be calibrated before performing this procedure. Refer to
Calibrating the Disinfectant Flowmeter
on page 28.
DO NOT use the disinfectant wand for showering. The disinfectant wand is to be used for disinfecting ONLY.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of
these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use cleaners
containing iodine, bromine, Betadine or methylene blue. The use of these types of cleaners will cause the
tub to stain. Never use cleaners containing bleach. Use of these types of cleaners will dry out the rubber
seals and gaskets and the tub will not function properly.
Disinfecting the Tub Using the Closed Loop System
NOTE:Forthisprocedure,refertoFIGURE 2.1.
NOTE:Ensurethatthewhirlpooljetsandwhirlpool inletfittingareclosedbeforestartingthispro ced ur e.
1. Cleantheentire tubsurface toremov evisibletissueresidueandfluid.
2. Dependingonthetypeofwhirlpooljetsthatareinyourtub,clo sealljetsbyperforming
oneofthe following:
•Turnthecenterofthejetclockwisetoafullstop.
•Pushthecenterofthejetin.
3. Closethewhirlpoolinletbyge ntly pullingthepluginthecenteroftheinletcoveroutw arduntilitstops.
4. Openaerato rbyrotatingaeratorknobtw oturnscounte r clockwise.
5. Open
disinfectantcabinetonthetopofthetub.
6. Ensurethat disinfectantsiphontubeisplacedin
disinfectantcontainer.
7. Holdthedisinfectantwandovertheinteriorofthetub.
8. Adjustwatertemperatureto105°F(40°C).
9. Turndisi nfecta ntOn/OffvalvetotheOnposition.
NOTE:Liquid will beginflowingfromthe disinfectantwandinto
the
tub.
10. Turntheadjustmentknobontheflowmeteruntilthe
floatingballrisestoalev elof35cc/min.
NOTE:At35cc/min.,disinfectantisbeingmixedwithwaterat
thepr operratio.
11. Continue tohold thedisinfectantwandoverthe
interiorofthetub.
NOTE:Whendisinfectanthastotallyfilledtheinteriorplumbing,
liquidwillbegintoflowfro mthedisinfectantwand.
NOTE:Invacarerecommendsthatyouthor oughlyscruball
interiorsurfacesofthetubwithdisinfectant.Theuseofalo ng
handledbrushwillmakethisoperationeasier.
12. Spray theinteriorsurfacesofthetub,includingthe
whirlpooljets,ov erflowfitting
andwhirlpoolinlet
fittingwithdisinfectant.Scruballinteriorsurfaces.
13. TurnthedisinfectantOn/Off valvetotheOffposition.
14. Ret urndisinfectantwandtoitsholder.
15. After scrubbingtheinteriorsurfaces,leav ethe
disinfectantintheloopandonthetubsurfacesforten
minutes.
16. Use theshower/massagewandonthepulsesetting
and
directthesprayintobothwhirlpooljetstobackflush
untilthewaterdisc hargingfromthewhirlpoolinletis
clear.
17. Rinse allinteriortubsurfaces.
18. ReverseSTEPS2and3toopenthewhirlpooljetsand
whirlpoolinlet.
FIGURE 2.1 Disinfecting the Tub Using the Closed Loop
System
Whirlpool Jets
Disinfectant
Wand
Overflow
Disinfectant
Siphon Tube
Whirlpool Inlet
Disinfectant Cabinet
Disinfectant
On/Off Valve
DETAIL “B” -
TUB DETAILS
Push In To
Close
Whirlpool Inlet
Plug
Ball
Turn Clockwise
to Close
Whirlpool Jet
DETAIL “A” -
WHIRLPOOL JET
AND INLET DETAILS
Flow meter
Adjustment Knob