Fluke PM6681 Outdoor Timer User Manual


 
9-58 Command Reference
:MEASure:MEMory<N>? PM6681
Memory Recall, Measure and Fetch Result
Use this command when you want to measure several parameters fast.
:MEAS:MEM1? recalls the contents of memory 1 and reads out the result,
:MEAS:MEM2? recalls the contents of memory two and reads out the result etc.
The equivalent command sequence is *RCL1;READ?
The allowed range for <N> is 1 to 9. Use the somewhat slower :MEAS:MEMory?
N command described below if you must use memories 10 to 19.
TIMING
Command
Data Format
ASCii REAL
:MEAS:MEM1?
7.9 ms 6.7 ms
:MEAS:MEM? 1
9.1 ms 8.0 ms
*RCL 1;READ?
10.1 ms 8.9 ms
Returned format:
<measurement result>¿
Complies to standards: SCPI 1991.0, confirmed
:MEASure:MEMory? PM6681
8 <N>
Memory Recall, Measure and Fetch Result
Same as above command but somewhat slower. Allows use of all memories (1 to
19).
Example: :MEAS:MEM 8 13
This example recalls the instrument setting in memory number 13, makes a meas
-
urement, and fetches the result.
Complies to standards: SCPI 1991.0, confirmed