Fluke 9011 Weather Radio User Manual


 
9011 Dual-Well Calibrator
User’s Guide
40
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point s[etpoint] s set: 9999.99 {C or
F}
set: 150.00 C
Set current set-point to n s[etpoint]=n s=450 Controller
Range
Read temperature unit u[nits] u u:{C or F} u:C
Set temperature units: u[nits]=c/f
CorF
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
u[nits]=f u=f
Read scan function sc[an] sc scan: {ON or OFF} scan: ON
Set scan function: sc[an]=on/of[f]
ON or OFF
Turn scan function on sc[an]=on sc=on
Turn scan function off sc[an]=of[f] sc=of
Read scan rate sr[ate] sr srat: 999.9 {C or
F}/min
srat: 10.0
C/min
Setscanrateto
n
degrees per
minute
sr[ate]=
n
sr=5 0.1 to 50°C
Read display temperature hold
mode
hm[ode] hm hmode: AUTO hmode: AUTO
Read hold status ho[ld] ho ho:{closed or
open}, 9999.9 {C or
F}
ho: open,
75.0 C
Read temperature t[emperature] t t: 9999.9 {C or F} t: 55.69 C
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 999.99999 pb: 15.9
Set proportional band to
n
pr[op-band]=n pr=8.83 Depends on
Configuration
Read cutout setting c[utout] c cu: 9999 {C or F} cu: 620 C, in
Read heater power
(duty cycle)
po[wer] po po: 9999.9 po: 1.0
Ramp and Soak Menu
Read number of programmable
set-points
pn pn pn: 9 pn: 2
Set number of programmable
set-points to
n
pn=
n
pn=4 2 to 8
Read programmable set-point
number
n
ps
n
ps3 ps
n
: 9999.99 {C or
F}
ps1: 50.00 C
Set programmable set-point num
-
ber
n
to
n
ps
n
=
n
ps3=50 1 to 8, Instru
-
ment Range
Table 2 Controller Communications Commands