Fluke 9011 Weather Radio User Manual


 
grounding plug for your protection against electrical shock hazards. It
must be plugged directly into a properly grounded three-prong receptacle.
The receptacle must be installed in accordance with local codes and ordi
-
nances. Consult a qualified electrician. DO NOT use an extension cord or
adapter plug.
If supplied with user accessible fuses, always replace the fuse with one of
the same rating, voltage, and type.
Always replace the power cord with an approved cord of the correct rat
-
ing and type.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE
INJURY or DEATH may result if personnel fail to observe safety precau
-
tions. Before working inside the equipment, turn power off and discon
-
nect power cord.
1.3.2
CAUTIONS
Always operate this instrument at room temperature between 41°F and
122°F (5°C to 50°C). Allow sufficient air circulation by leaving at least 6
inches (15 cm) of clearance around the instrument.
Component lifetime can be shortened by continuous high temperature op-
eration.
DO NOT apply any type of voltage to the display hold terminals. Apply-
ing a voltage to the terminals may cause damage to the controller.
DO NOT use fluids to clean out the well. Fluids could leak into electron-
ics and damage the instrument.
Never introduce any foreign material into the probe hole of the insert.
Fluids, etc. can leak into the instrument causing damage.
DO NOT change the values of the calibration constants from the factory
set values. The correct setting of these parameters is important to the
safety and proper operation of the calibrator.
DO NOT slam the probe sheath in to the well. This type of action can
cause a shock to the sensor and affect the calibration.
The instrument and any thermometer probes used with it are sensitive in
-
struments that can be easily damaged. Always handle these devices with
care. DO NOT allow them to be dropped, struck, stressed, or overheated.
The Factory Reset Sequence (see Section , Troubleshooting) should be
performed only by authorized personnel if no other action is successful in
correcting a malfunction. You must have a copy of the most recent Report
of Calibration to restore the calibration parameters.
DO NOT operate this instrument in an excessively wet, oily, dusty, or
dirty environment. Always keep the well and inserts clean and clear of
foreign material.
The dry-well is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with
5
1 Before You Start
Safety Information