Fluke 123 Telescope User Manual


 
Fluke 123/124
Users Manual
66
Accuracy:
@1 Hz to 1 MHz............................±(0.5% +2 counts)
@1 to 10 MHz ...............................±(1.0% +2 counts)
@10 to 50 MHz (Fluke 123) ..........±(2.5% +2 counts)
@10 to 70 MHz (Fluke 124) ..........±(2.5% +2 counts)
(50 MHz in Autorange)
Full Scale Reading..................................10 000 counts
Duty Cycle (DUTY)
Range...........................................................2% to 98%
Frequency Range in Continuous Autoset .....................
15 Hz (1 Hz) to 30 MHz
Accuracy (Logic or Pulse waveforms):
@1Hz to 1 MHz.............................±(0.5% +2 counts)
@1 MHz to 10 MHz .......................±(1.0% +2 counts)
Pulse Width (PULSE)
Frequency Range in Continuous Autoset .....................
15 Hz (1 Hz) to 30 MHz
Accuracy (Logic or Pulse waveforms):
@1 Hz to 1 MHz............................±(0.5% +2 counts)
@1 MHz to 10 MHz .......................±(1.0% +2 counts)
Full Scale Reading..................................1000 counts
Amperes (AMP) ....................with optional current probe
Ranges......... same as VDC, VAC, VAC+DC, or PEAK
Scale Factor...1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A
Accuracy .........same as VDC, VAC, VAC+DC, or PEAK
(add current probe uncertainty)
Temperature (TEMP) ... with optional temperature probe
Range........................................200 °C/div (200 °F/div)
Scale Factor ............................... 1 mV/°C and 1 mV/°F
Accuracy........... as VDC (add temp. probe uncertainty)
Decibel (dB)
0 dBV........................................................................ 1V
0 dBm (600 /50)............................................. 1 mW
referenced to 600 or 50
dB on...................................... VDC, VAC, or VAC+DC
Full Scale Reading .................................... 1000 counts
Crest Factor (CREST)
Range.................................................................1 to 10
Accuracy............................................... ±(5% +1 count)
Full Scale Reading ...................................... 100 counts
Phase
Modes...................................................... A to B, B to A
Range.................................................0 to 359 degrees
Accuracy up to 1 MHz ...................................2 degrees
Accuracy 1 MHz to 5 MHz.............................5 degrees
Resolution .......................................................1 degree