Do not insert metal objects into connectors.
Always use the Test Tool only in the manner
specified.
Max. Input Voltages
Input A and B directly...............................600 V CAT III
Input A and B via BB120.......................... 300 V CAT III
Input A and B via STL120 ........................600 V CAT III
Max. Floating Voltage
From any terminal to ground....................600 V CAT III
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50-60 Hz) for AC sinewave
applications and as Vdc for DC applications.
Overvoltage Category III refers to distribution level and
fixed installation circuits inside a building.
The terms ‘Isolated’ or ‘Electrically floating’ are used in
this manual to indicate a measurement in which the Test
Tool’s Shielded Banana inputs or banana jack is
connected to a voltage different from earth ground.
The isolated input connectors have no exposed metal and
are fully insulated to protect against electrical shock.
If Safety Features are Impaired
Use of the Test Tool in a manner not specified may
impair the protection provided by the equipment.
Before use, inspect the test leads for mechanical damage
and replace damaged test leads!
Whenever it is likely that safety has been impaired, the
Test Tool must be turned off and disconnected from the
line power. The matter should then be referred to qualified
personnel. Safety is likely to be impaired if, for example,
the Test Tool fails to perform the intended measurements
or shows visible damage.