Fluke 123 Telescope User Manual


 
Maintaining the Test Tool
Using and Adjusting 10:1 Scope Probes
2
47
Note
Ensure that the battery pack is placed in the
battery compartment as shown in Figure 2-2.
For Fluke 123 use the Fluke BP120 Ni-Cd
(standard) or BP130 Ni-MH (extended operating
time) battery pack.
For Fluke 124 it is recommended to use the
Fluke BP130 Ni-MH battery pack .
8. Reinstall the battery cover and secure the screw.
Using and Adjusting 10:1 Scope Probes
Note
The 10:1 voltage probe that is supplied with
Fluke 124 is always adjusted correctly to the
Test Tool and needs no further adjustment.
You need to adjust other scope probes (e.g. VPS100,
optional) for optimal response.
Warning
To avoid electrical shock use the BB120
Banana-to-BNC adapter (delivered with the
test tool) to connect a 10:1 scope probe to
the input of the test tool.
To adjust probes, do the following:
Connect the 10:1 scope probe from the gray input B
jack to the red input A jack. Use the red 4-mm
banana adapter (delivered with the probe) and the
banana-to-BNC adapter (BB120). See Figure 2-3.
BB120
Figure 2-3. Adjusting Scope Probes