276 Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Creating and Downloading Waveform Files
Troubleshooting Waveform Files
Troubleshooting Waveform Files
Symptom Possible Cause
ERROR 224, Text file busy
Attempting to download a waveform that has the same name as the waveform
currently being played by the signal generator.
To solve the problem, either change the name of the waveform being downloaded
or turn off the ARB.
ERROR 628, DAC over range The amplitude of the signal exceeds the DAC input range. The typical causes are
unforeseen overshoot (DAC values within range) or the input values exceed the
DAC range.
To solve the problem, scale or reduce the DAC input values. For more information,
see “DAC Input Values” on page 188.
On the Agilent MXG, this error can occur if an encrypted file (.SECUREWAVE) is
being downloaded to the signal generator from a PC or USB Media with a different
suffix (i.e. not .SECUREWAVE).
To solve the problem, use the Use as or Copy File to Instrument softkey menus to
download the encrypted file to the instrument. For more information, see
“Encrypted I/Q Files and the Securewave Directory (Agilent MXG)” on page 205.
ERROR 629, File format invalid The signal generator requires a minimum of 60 samples to build a waveform and
the same number of I and Q data points.
ERROR -321, Out of memory
There is not enough space in the ARB memory for the waveform file being
downloaded.
To solve the problem, either reduce the file size of the waveform file or delete
unnecessary files from ARB memory. Refer to “Waveform Memory” on page 198.
No RF Output The marker RF blanking function may be active. To check for and turn RF blanking
off, refer to “Configuring the Pulse/RF Blank (Agilent MXG)” on page 277 and
“Configuring the Pulse/RF Blank (ESG/PSG)” on page 277. This problem occurs
when the file header contains unspecified settings and a previously played
waveform used the marker RF blanking function.
For more information on the marker functions, see the User’s Guide.
Undesired output signal Check for the following:
The data was downloaded in little endian order. See “Little Endian and Big
Endian (Byte Order)” on page 186 for more information.
The waveform contains an odd number of samples. An odd number of samples
can cause waveform discontinuity. See “Waveform Phase Continuity” on
page 195 for more information.