214 Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Creating and Downloading Waveform Files
Creating Waveform Data
7–11 Create a loop to do the following:
Generate and scale the I data (DAC values). This example uses a simple sine equation, where 2*3.14
equals one waveform cycle. Change the equation to fit your application.
The array pointer, index, increments from 0–499, creating 500 I data points over one period of the
sine waveform.
Set the scale of the DAC values in the range of 32768 to 32767, where the values 32768 and 32767
equal full scale negative and positive respectively. This example uses 23000 as the multiplier,
resulting in approximately 70% scaling. For more information on scaling, see “Scaling DAC Values” on
page 189.
NOTE The signal generator comes from the factory with I/Q scaling set to 70%. If you reduce the DAC
input values, ensure that you set the signal generator scaling (:RADio:ARB:RSCaling) to an
appropriate setting that accounts for the reduced values.
Generate and scale the Q data (DAC value). This example uses a simple cosine equation, where 2*3.14
equals one waveform cycle. Change the equation to fit your application.
The array pointer, index, increments from 0–499, creating 500 Q data points over one period of the
cosine waveform.
Set the scale of the DAC values in the range of 32767 to 32768, where the values 32767 and 32768
equal full scale negative and positive respectively. This example uses 23000 as the multiplier,
resulting in approximately 70% scaling. For more information on scaling, see “Scaling DAC Values” on
page 189.
Line Code Description—Create I and Q data