2
If specified, standard fixed width slits 1m high are supplied to customer’s specified width.
Movable Diverter Mirror (optional): A movable diverter mirror is provided to divert the beam
from the V slit position to the straight through position. A knob in the instrument cover allows
the mirror to be moved under vacuum conditions
Slit Chamber Isolation Valve (optional): Vacuum valves are provided to isolate the slit
chambers form the main instrument chamber. Ports (1/8 NPT) are provided in the slit
chambers for gauge and roughing accessories
Air Inlet Valve (optional): An air inlet valve is located in the instrument housing to vent the
instrument to atmospheric pressure.