9-48 Command Reference
:INPut :LEVel PM6685
8 «<Decimal data>|MAX|MIN»
Waveform compensation
The three-position waveform compensation on the front panel is not available
from the bus. Instead, you can set the trigger level, that is, the level on which the
hysteresis band is centered. How to set the trigger level depends on the duty cycle
and the peak-to-peak voltage of the signal.
Trigger level V Duty factor
pp
=-*( . )05
This setting has no effect unless autosensitivity is turned off, see the following
page.
Parameters
<Decimal data> is a number between approximately –37.7 V and +37.7 V.
MAX gives +37.7 ... V and MIN gives –37.7... V
Note that the 8 :INP:LEV command is coupled with the :INP:HYST command.
See page 9-45.
Returned format: <Decimal data>¿
Example:
SEND® :INP:LEV 8 3.75;LEV:AUTO 8 0
This example sets the trigger level to 3.75 V and switches off auto trigger level.
*RST condition 0 (but controlled by Autotrigger since AUTO is on after *RST)