Fluke PM6681R Outdoor Timer User Manual


 
Command Reference 9-45
:INPut :HYSTeresisPM6685
8 «<Decimal data>|MAX |MIN»
Sensitivity
The sensitivity setting on the front panel is called HYSTeresis from the bus. The
range is 27.12 mV to 75.4 V. This setting has no effect unless autosensitivity is
turned off, see the following page.
Note that the sensitivity setting is coupled with the hysteresis setting according to
the formula:
Trigger Level Hysteresis+
<
2
377057. K
Parameters: <Decimal data> is a number between 27.12E-3 and 75.4.
MAX gives +75.4 V, MIN gives +2.7 mV
When using MAX as data, the counter always tries to set the hysteresis to +75.4 V.
Unless the Trigger level is set to 0, this setting is impossible, and the counter will
return an error message.
Returned format: <Decimal data>
Example:
SEND
:INP:HYST 8 0.5;HYST:AUTO 8 0
This example sets the sensitivity to 0.5 V and switches off autosensitivity.
*RST condition 0.65 V (but controlled by Autotrigger since AUTO is on after *RST)
:INPut :FILTerPM6680B/81/85
8 <Boolean>
Low Pass Filter
Switches on or off the low pass filter on input 1 (A). It has a cutoff frequency of 100
kHz.
Parameters:
<Boolean> is (1 / ON|0/OFF)
Returned format: 1|0¿
*RST condition OFF
Complies to standards: SCPI 1991.0, confirmed.