PRBS
See “Data Pattern” on page 79 for for the timing diagram.
Characteristics A PRBS signal can be either generated continuously or can be
started manually or by the arming source.
The PRBS polynome is selectable from 2
5
– 1 … 2
31
– 1.
The instrument mode is Pulse/Pattern.
The pattern mode is PRBS.
You can select between RZ, NRZ and R1 signal modes for each
output.
In R1 and RZ modes, the pulse width can be selected as either width
or duty cycle.
You can select between normal and inverted signals for each output.
For the clock source, you can select from:
Internal (YIG Oscillator)
External signal at Clock Input
External 10 MHz Reference at Clock Input
For more information about the clock sources, see “Clock Sources”
on page 73.
The Trigger Out marks every start of the repetitive PRBS, as long as
no frequency divider has been applied.
See “Trigger Out” on page 75.