Agilent Technologies 54621D Sprinkler User Manual


 
3-20
Agilent 54622A/22D/24A Performance Test Record
Agilent 54622A/54622D/54624A
Serial No. ______________________________________ Test by _____________________________
Test Interval ____________________________________ Work Order No. ______________________
Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8
Accuracy Test
5 V - 250 mV 4.750 V ________ ________
(100 mV + 3% of
threshold setting)
5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________
0 V - 100 mV -100 mV ________ ________
0 V + 100 mV 100 mV ________ ________
Voltage Measurement Accuracy
Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________
2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________
1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________
500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________
200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________
100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________
50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________
20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________
10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________
5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________
2 mV/Div 14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________
1 mV/Div 7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
Bandwidth Test Limits Channel 1 Channel 2 Channel 3 Channel 4
3 dB at 100 MHz ________ ________ ________ ________
Horizontal t and 1/t Accuracy
Generator Setting Test Limits Results
Frequency 10 kHz 9.98 kHz to 10.02 kHz ________
Period 100
µs 99.79 µs to 100.2 µs ________
Frequency 1 MHz 997.9 kHz to 1.002 MHz ________
Period 1
µs 997.9 ns to 1.002 µs ________
Frequency 100 MHz 99.10 MHz to 100.9 MHz ________
Period 10 ns 9.91 ns to 10.09 ns ________
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3 Channel 4
Internal trigger 25 MHz at 0.35 division ________ ________ ________ ________
100 MHz at 0.35 division ________ ________ ________ _______
Ext
External trigger 25 MHz at <75 mV
p-p
________
100 MHz at <150 mV
p-p
________
service.book Page 20 Wednesday, December 18, 2002 8:35 AM