Tektronix AFG3000 Series Portable Generator User Manual


 
Getting Started
16 AFG3000 Series Quick Start User Manual
6. Select the downloaded firmware file by
rotating the general purpose knob, and
then push the Execute bezel button.
The firmware file name is as follows:
tekafgtb-1.x.x.tfb
7. The instrument asks you "Are you sure
you want to update firmware?". Push
OK.
8. The instrument displays "Do not power
off until the process completes
." The
clock symbol at the top right of the
screen indicates the update process is
in progress.
CAUTION. A firmware update usually
takes approximately two minutes. Do not
remove the USB memory during the
update process.
CAUTION. If you accidentally removed the
USB memory during the update process,
do not power off the instrument. Repeat
the installation process from step 3.
9. Wait until the instrument displays
"Operation completed".
10. Push OK.
CAUTION. If “Operation completed” is not
displayed, do not power off the instrument.
Repeat the installation process from step 2
using a different type of USB memory.