National Instruments NI PXI-5422 Portable Generator User Manual


 
© National Instruments Corporation 27 NI PXI-5422 Calibration Procedure
21 0.018500 0.009250 – 0.009250 ±0.000532 ±0.000579
22 0.013000 0.006500 – 0.006500 ±0.000523 ±0.000555
23 0.009000 0.004500 – 0.004500 ±0.000516 ±0.000538
24 0.006500 0.003250 – 0.003250 ±0.000511 ±0.000528
1
Error Positive Offset Value = (Measured Positive Offset Value) – (Ideal Positive Offset Value)
2
Error Negative Offset Value = (Measured Negative Offset Value) – (Ideal Negative Offset Value)
Table 5. Values for Verifying the Offset of the Main Analog Path (Continued)
Iteration Gain
Ideal
Positive
Offset
(Volts)
Ideal
Negative
Offset
(Volts)
Measured
Positive
Offset
(Volts)
Measured
Negative
Offset
(Volts)
Error
Positive
Offset
(Volts)
1
Error
Negative
Offset
(Volts)
2
Calibration
Test Limit
(Volts)
Published
Spec
(Volts)