External Calibration Requirements
This section describes the test equipment and test conditions required for
calibration.
Test Equipment
External calibration requires different equipment for each applicable
specification. Refer to Table 2 for a list of equipment.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
Keep connections to the NI PXI-5422 short.
Keep relative humidity below 80%.
Maintain a temperature between 18 °C and 28 °C.
Observe the 15-minute warm-up time.
Table 2. Equipment Required for Calibrating the NI PXI-5422
Instrument
Applicable
Specification
Minimum
Specifications
Recommended
Instrument
Digital multimeter
(DMM)
DC gain and offset DC accuracy ±50 ppm
Resolution 1 µV
NI PXI-4070
Agilent HP 34401A
Keithley 2000
Male banana to
female BNC adapter
Male BNC to
female SMB cable
50 Ω, RG-223
Spectrum analyzer or
frequency meter
Frequency accuracy Ability to measure
10 MHz or higher
sine waves
Frequency accuracy
to ±500 ppb
NI PXI-5660
Agilent HP 8560E
Agilent HP 53131A
or HP 53132A
with timebase option
001, 010, or 012
Male BNC to
female SMB cable
50 Ω, RG-223