© National Instruments Corporation 23 NI PXI-5422 Calibration Procedure
21 0.018500 0.018500 – 0.018500 ±0.000530 ±0.000574
22 0.013000 0.013000 – 0.013000 ±0.000521 ±0.000552
23 0.009000 0.009000 – 0.009000 ±0.000514 ±0.000536
24 0.006500 0.006500 – 0.006500 ±0.000510 ±0.000526
1
Error Positive Full-Scale Value = (Measured Positive Full-Scale Value) – (Ideal Positive Full-Scale Value)
2
Error Negative Full-Scale Value = (Measured Negative Full-Scale Value) – (Ideal Negative Full-Scale Value)
Table 4. Values for Verifying the Gain of the Main Analog Path (Continued)
Iteration Gain
Ideal
Positive
Full-Scale
(Volts)
Ideal Negative
Full-Scale
(Volts)
Measured
Positive
Full-Scale
(Volts)
Measured
Negative
Full-Scale
(Volts)
Error
Positive
Full-Scale
(Volts)
1
Error
Negative
Full-Scale
(Volts)
2
Calibration
Test Limit
(Volts)
Published
Spec
(Volts)