© National Instruments Corporation 21 NI PXI-5422 Calibration Procedure
12. Call niFgen_SetAttributeViReal64 to set the offset (niFgen
property node: Arbitrary Waveform Output»Arbitrary Waveform
Offset) using the following parameters:
• vi: The session handle returned from
niFgen_init
• channelName: "0"
• attributeID: NIFGEN_ATTR_ARB_OFFSET
• value: 0
13. Repeat steps 14 through 24 for each of the 24 iterations listed in
Table 4, changing the Gain value for each iteration. You can use
Table 4 to record the results of these steps.