Anritsu MW9076 Telescope User Manual


 
Model MW9076B MW9076C MW9076B1 MW9076J MW9076K MW9076D1
15
1 At 25°C, pulse width: 1 µs
2 For GI fiber (core diameter: 62.5 µm ±3.0 nm, NA: 0.275 ±0.015, transmission loss: 3.2/0.9 dB/km
(wavelength: 0.85/1.3 µm). At measurement of 50/125 µm GI fiber, the dynamic range drops by about
3.0 dB.
3 At 25°C, pulse width: SM 20 µs, GI 100 ns (0.85 µm), 1 µs (1.3 µm)
4 Pulse width: 10 ns, return loss: SM 40 dB, GI 30 dB, deviation: ±0.1 dB (Refer to the figure right.)
5 Pulse width: 10 ns (Refer to the figure right.)
6 Either value is automatically selected in each mode, depending on the distance range.
7 Automatic measurement is a supporting function which enables to operate easier, it doesn’t assure
results. As there is a case of miss detection, please check a waveform data, either.
8 At quick mode
9 Compared value with internal wavelength data at chromatic dispersion measurement
10 Measured with 25 km of 1.3 µm zero-dispersion fiber (ITU-T G.652) at 1550 nm.
Not an error from absolute value but repeatability of measured results. Contact Anritsu Corporation in
case of measuring ITU-T G.655 fiber.
11 At back light low brightness, measurement not executed.
12 Dropped on the floor of plywood thickness 5 cm fixed by concrete. Not applicable to the MW9076D1.
l2
1.5 dB
0.1 dB
l1: Fresnel reflection
l2: Back-scattered light
l1
Operating temperature and humidity: –10° to 40°C, 85% (no condensation)
Storage temperature and humidity: –20° to 60°C, 85%
Vibration: Conforming to MIL-T-28800E Class 3
Environmental condition
Shock: 76 cm height, 6 surfaces, 8 corners
12
Dust-proofing: MIL-T-28800E
Drip-proofing: MIL-T-28800E
EN61326: 1997/A1, 1998 (Class A)
EMC EN61000-3-2: 1995/A2, 1998 (Class D)
EN61326: 1997/A1, 1998 (Annex A)
LVD EN610101-1: 1993/A2, 1995 (Installation Category ΙΙ, Pollution Degree 2)