Agilent Technologies 54754a Sprinkler User Manual


 
Command
:STORe:WAVEform {CHANnel<number> |
FUNCtion<number> | WMEMory<number> | HISTogram |
FFT | RESPonse<number>},{WMEMory<number>}
Copies a channel, function, stored waveform, histogram, TDR response, or
FFT to a waveform memory. The first parameters specifies the source of the
copy and can be any channel, function, waveform memory, histogram, TDR
response, or the FFT. The second parameter is the destination of the copy,
and can be any waveform memory.
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For responses: 1, 2, 3, or 4.
Command
:VIEW {CHANnel<number> | WMEMory<number> |
FUNCtion<number> | FFT | RESPonse<number> |
HISTogram | PMEMory1}
Turns on a channel, function, pixel memory, waveform memory, TDR
response, histogram, or FFT. Use the BLANk command to turn these off.
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4. For
pixel memories (PMEMory): 1.
Root Level Commands
6