Command
:TDR{2 | 4}:RATE[?] {AUTO | <rate>}
Sets the period of the TDR pulse generator. You should usually leave this set
to AUTO unless you need to define a specific rate. In AUTO, the instrument
will attempt to keep subsequent periods off screen when the timebase is
changed.
<rate>
Period to which you want to set the generator, in Hertz. You can add a suffix
to indicate that the rate is in Hertz (HZ, KHZ, MHZ, and so on).
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}[?] {OFF |
DIFFerential | COMMonmode | NORMalize}
Turns on or off the response of a TDR or TDT stimulus.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:HORizontal[?]
{TSOurce, MANual}
Specifies whether the TDR/TDT response should track the source channel’s
horizontal scale (TSOurce), or use a user-defined scale specified with the
HORizontal:POSItion and HORizontal:RANGe commands (MANual). TSOurce
is the usual setting.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4):
HORizontal:POSition[?] <position>
Specifies the horizontal position of the TDR/TDT response when horizontal
tracking is set to manual. The position is always referenced to center screen.
<position>
Offset from the center of the screen, in seconds.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:
HORizontal:RANGe[?] <range>
Specifies the range of the TDR/TDT response when the horizontal tracking is
set to manual.
<range>
Horizontal range in seconds.
TDR Commands