Agilent Technologies 54621A Sprinkler User Manual


 
4-46
MegaZoom Concepts and Oscilloscope Operation
To define a new label
Label List Management
When you press the Library softkey, you will see a list of the last 75 labels used. The
list does not save duplicate labels. Labels can end in any number of trailing digits.
As long as the base string is the same as an existing label in the library, the new label
will not be put in the library. For example, if label A0 is in the library and you make a
new label called A12345, the new label is not added to the library.
When you save a new user-defined label, the new label will replace the oldest label
in the list. Oldest is defined as the longest time since the label was last assigned to
a channel. Any time you assign any label to a channel, that label will move to the
newest in the list. Thus, after you use the label list for a while, your labels will
predominate, making it easier to customize the instrument display for your needs.
When you reset the label library list (see next topic), all of your custom labels will be
deleted, and the label list will be returned to its factory configuration.