27 Basic setups
Meter 1/Meter 2 Parameters (i.e. response time, weighting)
Meter 1/Meter 2 Parameters (i.e. response time,
weighting)
Setting up Meter 1 and Meter 2 parameter settings are changed through the
softkeys and the Measure Set screen of the SoundPro.
Setting up Meter 1 and/or Meter 2 parameters
1. Ensure you are at the start screen.
• If turned on, press On/Off/Esc repeatedly until you return to the
start screen (see Figure 3-3).
2. Press Up/Down to highlight Setup and then press Enter .
3. Select Meter Set by pressing Up/Down to highlight and then press
Enter .
4. To set the response time for Meter 1, press the second softkey repeatedly
until the either Fast (F), Slow (S), or Impulse (I) is selected.
• An underline denotes which feature is selected.
5. To set the weighting for Meter 1, press the third softkey repeatedly until
either A, C, Z, of F is underlined (or selected).
• To set Meter 2 response time and weighting, press 4
th
softkey until
Meter 2 is selected. Then repeat steps 4 & 5.
6. To change the Meter parameters, press Up/Down to select one of
the Meter Set fields.
7. To change any of the values, press Enter to move to the data fields.
• Continue to press Enter until the specified setting is selected. Press
the Left arrow to return to the Meter Set fields.
Figure 3-4: Meter 1 and Meter 2 parameters
Press up/down
arrows to move
between the
parameters
Press Enter
repeatedly to
toggle through the
selections
2
ND
softkey– Response time
3
RD
softkey– Weighting
4
th
softkey– Meter 1/Meter 2