MODEL 2230R, Process Hydrogen Analyzer
Page 26
21 Settle time: 30 min
(Y/N)?
Typey”
22 Apply 30.000%H2:
Ready (Y/N)?
Typey”
Streaming data…
23 Taking Average…
cap=0.00000
res=x.xxxxx
Calibration Gas #2 finished.
Calibration complete.
4) Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of
Calibration, conduct the Verification sequence as described in the previous section to
verify that the calibration process was executed correctly.