MODEL 2230R, Process Hydrogen Analyzer
Page 25
FIELD CALIBRATION STEPS
Step Display User response
1 H2scan: Type f to run field calibration
2 Ready to Calibrate
(Y/N)?
Typey”
3 Pressure is 0.9690
atm (Y/N)?
Typey”
Please note that the firmware has remembered the
atmospheric pressure from the preceding Verification
procedure. Type n if a change is required and enter the
correct pressure.
4 Save as default
(Y/N)?
Type n
5 Gas1 is 0%H2
(Y/N)?
Type n
6 Gas1 for res (Y/N)? Typey”
7 Cal Gas: 2.000%H2
(Y/N)?
Type n
8 Enter gas: Type 1 for gas #1, 1% hydrogen by volume
9 Cal Gas: 1.000%H2
(Y/N)?
Typey”
10 Settle time: 3 min
(Y/N)? n
Type n
11 Enter time: Type 30 for a 30 minute duration for gas #1
12 Settle time: 30 min
(Y/N)?
Typey”
13 Apply 1.000%H2:
Ready (Y/N)? y
Typey”
Streaming data…
14 Taking Average…
cap=0.00000
res=x.xxxxx
Calibration Gas #1 finished
15 Gas2 for res (Y/N)? Typey”
16 Cal Gas:
100.000%H2 (Y/N)?
Type n
17 Enter gas: Type 30 for gas #2, 30% hydrogen by volume
18 Cal Gas: 30.000%H2
(Y/N)?
Typey”
19 Settle time: 1 min
(Y/N)?
Type n
20 Enter time: Type 30 for a 30 minute duration for gas #2