Agilent Technologies E8257D/67D Portable Generator User Manual


 
30 Chapter 3
Operation Verification
Checking the Output Power
Table 3-5 Leveled Output Power Limits, E8257D cont.
E8257D
Option 550 Option 567
Frequency Amplitude Limits Frequency Amplitude Limits
Option dBm dB Option dBm dB
250 kHz None 9 ±0.6 250 kHz None 9 ±0.6
1EA 15 ±0.6 1EA 15 ±0.6
1E1 7 ±0.6 1E1 7 ±0.6
1E1 & 1EA 14 ±0.8 1E1 & 1EA 14 ±0.8
3.2 GHz None 9 ±0.8 3.2 GHz None 9 ±0.8
1EA 15 ±0.8 1EA 15 ±0.8
1E1 7 ±0.8 1E1 7 ±0.8
1E1 & 1EA 14 ±0.8 1E1 & 1EA 14 ±0.8
20 GHz None 9 ±0.9 20 GHz None 9 ±0.9
1EA 18 ±1.0 1EA 18 ±1.0
1E1 7 ±0.9 1E1 7 ±0.9
1E1 & 1EA 16 ±1.0 1E1 & 1EA 16 ±1.0
40 GHz None 9 ±0.9 40 GHz None 9 ±0.9
1EA 14 ±1.0 1EA 14 ±1.0
1E1 7 ±0.9 1E1 7 ±0.9
1E1 & 1EA 12 ±1.0 1E1 & 1EA 12 ±1.0
50 GHz None 5 ±1.2 50 GHz None 5 ±1.2
1EA 14 ±1.2 1EA 14 ±1.2
1E1 3 ±1.2 1E1 3 ±1.2
1E1 & 1EA 10 ±1.2 1E1 & 1EA 10 ±1.2